Specifications:
Left-handed equation、Right type
[Product performance, structural composition, and scope of application]
Composed of a pair of overlapping tail blades. Made of 05Cr17Ni4Cu4Nb material specified in GB/T 1220-2007 or TC4 material specified in GB/T 2965-2007. Non sterile provision. The product has good corrosion resistance, a smooth surface, and is sterilized by the user according to the instructions before use. Used for gripping eye tissue, intraocular foreign bodies, and instruments.
[Contraindications]
None
[ Precautions]
During high-pressure steam sterilization, do not close the strabismus forceps, otherwise it will affect the sterilization effect. In addition, during high-pressure steam sterilization, thermal expansion can also cause the tail rivet to break. Used by professional medical personnel.
[Warning]
1. It is strictly prohibited to use strabismus forceps when they deform, break, or rust;
2. Non professionals are strictly prohibited from repairing deformed strabismus forceps.
[Instructions for Use]
1. High temperature and high pressure sterilization is required before use of the product;
2. After use, the strabismus forceps should be cleaned promptly.
[Product maintenance and upkeep methods]
1. After the surgery, the remaining stains on the strabismus forceps should be washed clean and dried for later use;
2. The strabismus forceps should be stored in a box with a soft material cushion to prevent collision between the strabismus forceps and damage to the inner surface of the strabismus forceps head.
[Storage]
Strabismus forceps should be stored indoors with a relative humidity of no more than 80%, no corrosive gases, and good ventilation.